Vacuum decay leak testing for syringes and cartridges

A non-destructive, deterministic leak test developed and validated for your prefilled syringe or cartridge, including containers removed from assembled devices.

How vacuum decay works

The sealed package is placed in a closed test chamber. The chamber is evacuated and then monitored: if the package leaks, gas or liquid escaping from it raises the chamber pressure.

Vacuum decay is one of the deterministic methods described in USP <1207>, and ASTM F2338 is the standard test method for it. It measures a physical change, the pressure in the chamber around the package, and gives a quantitative result rather than relying on a visual reading. The test is non-destructive.

What the method can detect depends on the package: the container, the fill volume, the headspace and the stopper position. Kiln™ runs vacuum decay on a PTI-325 vacuum decay leak tester with tooling made for your container, and sets the test parameters and the detection limit for your package with positive controls made in your own container.

123
  1. 1Test chamber, with tooling made for your container
  2. 2Seal interfaces, such as the stopper-to-barrel seal, where a leak path would open into the chamber
  3. 3The filled syringe or cartridge under test
The sealed package sits in a closed chamber. After the chamber is evacuated, a rise in chamber pressure indicates a leak.

What we do

We confirm that vacuum decay suits your package, build positive controls in your container, and develop and validate the method. Then we run it, or hand it over.

Method development and validation

What turns a vacuum decay test into a validated method for your product.

Feasibility and tooling

We first confirm that your package is compatible with vacuum decay, and fit test tooling to your syringe or cartridge.

  • Feasibility
  • Custom tooling

Positive and negative controls

Positive controls are made in your own container: laser-drilled defects of varied size and location in the glass, wire-channel defects at the stopper, and gross defects such as torn or punctured stoppers. Negative controls come from normal processing.

  • Laser-drilled defects
  • Wire-channel defects

Method development

Test parameters and pass/fail limits are set with units filled with placebo or surrogate solution, so drug product is saved. Trials with positive controls establish the detection limit, and repeatability and robustness are checked before validation.

  • Pass/fail limits
  • Detection limit

Validation

Multiple test series across operators and days, per USP <1225> and USP <1207>. Acceptance is plain: every negative control passes and every positive control fails. You receive the protocol, the report and a final test method.

  • USP <1225>
  • Protocol and report

Using the method

A validated method can be reused at each stage of the product life cycle.

Stability, transit and air shipment

The validated method is applied at stability timepoints, after distribution simulation, and after simulated air shipment, where a headspace bubble expanding at altitude can push the stopper.

  • Stability
  • Post-transit CCI

CCI of assembled devices

Where a device blocks access to the container, we develop and validate a disassembly method that frees the syringe or cartridge without damaging the seals, then test the container.

  • Device disassembly
  • Autoinjectors

Choosing between methods

Vacuum decay is one of five CCI methods we run. Where another method fits the package better, such as HVLD for liquid-filled packages, we will say so during scoping.

  • Method selection

How we work

Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.

  1. Scope

    We review the package, the product and the stage of your program, and recommend a method and a sample plan.

  2. Develop

    Tooling is fitted, positive controls are made in your container, and test parameters are optimized with placebo or surrogate.

  3. Validate

    Multiple test series across operators and days. Every negative passes, every positive fails, and the detection limit is documented.

  4. Test

    Samples from development, stability, transit studies or production lots are tested against the validated method.

  5. Support

    The method stays available for later timepoints, investigations and lot release, or transfers to your lab.

Deliverables

Deliverables are defined in the proposal before work starts. A typical engagement includes the following.

Feasibility resultWhether vacuum decay suits your package, and why
Positive controlsDefects of known size made in your own container and closure
Validated methodValidation protocol, report and final test method
Test reportsResults for each sample set, with the underlying data

Related services

Container closure integrity testingAll CCI methods, positive controls and validation to USP <1207>
High voltage leak detectionNon-destructive HVLD for liquid-filled syringes and cartridges
Helium leak testingInherent integrity of each closure interface per ASTM F2391
Headspace analysisLaser-based headspace gas analysis of the sealed container
Dye ingress testingA probabilistic method for continuity with existing data

Questions we hear often

If yours isn't here, it's a good thing to bring to a consultation.

Is vacuum decay destructive?

No. The package is tested in a closed chamber and is not opened or altered by the test.

How small a leak can vacuum decay detect?

It depends on the package and the test parameters. The detection limit is established for your container with positive controls of known size during method development, and documented in the validation.

Should we use vacuum decay or HVLD?

It depends on the package and the product. HVLD applies to liquid-filled packages and can detect defects down to 1 to 2 microns. Vacuum decay measures a pressure change in a chamber around the package. We recommend a method during scoping based on your container, your fill and the stage of your program.

Can you test the syringe inside an autoinjector?

Yes. We develop and validate a disassembly method that frees the primary container from the device without disturbing the seals, then test the container.

What happens if our package changes?

A change to the container, closure, fill volume or stopper position can change what the method detects, so the method may need to be requalified. We will tell you during scoping which changes would trigger that.

Need a vacuum decay method?

Tell us about the package and the product. We'll set up a call to talk through method options, samples and timing.

Prefer email? Write to info@kiln-lab.com or call +1 (866) 350-8240.