Container closure integrity testing (USP <1207>, ASTM F2391)

Leak test methods developed and validated for your specific package: prefilled syringes, cartridges and the containers inside assembled devices.

What USP <1207> asks for

USP <1207> is the general chapter on package integrity evaluation for sterile products. It covers how to choose a leak test method, how to validate it, and how to relate its sensitivity to the leakage the product can tolerate.

A sterile injectable has no second barrier, so integrity has to be shown across the product life cycle, not once. The chapter favors deterministic methods, which measure a physical property and give a quantitative, repeatable result, over probabilistic ones such as dye ingress. ASTM F2391 is the standard method for measuring package and seal integrity with helium as a tracer gas, and is the usual way to establish the inherent integrity of a closure during development.

A CCI method only works for the package it was built for. Fill volume, headspace, stopper position, the product and each seal interface all change what a method can detect, so a change to the package can mean requalifying the method. We build the method around your container from the start.

123
  1. 1Needle-side seal: needle shield or tip cap against the syringe tip
  2. 2Stopper-to-barrel seal, along the full stopper contact
  3. 3Container wall: where laser-drilled positive-control defects are made
Each seal interface of a prefilled syringe is isolated and evaluated on its own, and positive controls are made where a real leak would occur.

What we do

We select the method with you, build positive controls for your package, and develop and validate the method. Then we run it, or hand it over.

Leak test methods

Deterministic methods first, with dye ingress available where continuity with older data is needed.

Helium leak testing

Each closure interface, such as the stopper and the tip or crimp seal, is isolated with custom fixturing and its helium leak rate measured per ASTM F2391. This establishes the inherent integrity of the closure and compares components and configurations.

  • ASTM F2391
  • Inherent integrity

High voltage leak detection

The filled unit rotates while electrodes scan its outer contours. A breach shows as a change in resistance and current. High voltage leak detection can detect defects down to 1 to 2 microns. The test is non-destructive and needs little operator handling, which makes it practical for stability and lot testing.

  • HVLD
  • Non-destructive

Vacuum decay

A non-destructive method that detects a leak from a pressure change in a test chamber around the package, with tooling made for your container.

  • Vacuum decay
  • Non-destructive

Dye ingress

A probabilistic method, run with defined challenge conditions and positive controls, for programs that need continuity with existing data.

  • Probabilistic

Method development and use

What turns a leak test into a validated method for your product.

Positive and negative controls

Positive controls are made in your own container: laser-drilled defects of varied size and location in the glass, wire-channel defects at the stopper, and gross defects such as torn or punctured stoppers. Negative controls come from normal processing. Trials with both establish the detection limit.

  • Laser-drilled defects
  • Detection limit

Validation

Three test series by two operators on different days, per USP <1225>. Acceptance is plain: every negative control passes and every positive control fails. You receive the protocol, the report and a final test method.

  • USP <1225>
  • Protocol and report

CCI of assembled devices

Where a device blocks access to the container, we develop and validate a disassembly method that frees the syringe or cartridge without damaging the seals, then test the container.

  • Device disassembly
  • Autoinjectors

Stability, transit and air shipment

The validated method is applied at stability timepoints, after distribution simulation, and after simulated air shipment, where a headspace bubble expanding at altitude can push the stopper.

  • Stability
  • Post-transit CCI

How we work

Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.

  1. Scope

    We review the package, the product and the stage of your program, and recommend a method and a sample plan.

  2. Develop

    Positive controls are made in your container, and method parameters are optimized with placebo or surrogate so drug product is saved.

  3. Validate

    Three test series, two operators, different days. Every negative passes, every positive fails, and the detection limit is documented.

  4. Test

    Samples from development, stability, transit studies or production lots are tested against the validated method.

  5. Support

    The method stays available for later timepoints, investigations and lot release, or transfers to your lab.

Deliverables

Deliverables are defined in the proposal before work starts. A typical engagement includes the following.

Method recommendationWhich leak test fits your package, and why
Positive controlsDefects of known size made in your own container and closure
Validated methodValidation protocol, report and final test method
Test reportsResults for each sample set, with the underlying data

Related services

ISO 11040 testingComponent and finished-syringe testing for prefilled syringes and cartridges
Filling and surrogate solutionsSmall-scale filling at nominal and process extremes, with matched surrogates
Device assembly and packagingDevelopment builds with force and torque monitored on every unit
ISO 11608 testingDesign verification for pens, autoinjectors and other needle-based injection systems
Devices we supportPrefilled syringes, cartridges, autoinjectors, pens, wearables and more

Questions we hear often

If yours isn't here, it's a good thing to bring to a consultation.

Which CCI method is right for my package?

It depends on the package, the product and the leak size you need to detect. The methods we run are high voltage leak detection (HVLD), helium leak, vacuum decay and dye ingress. HVLD can detect defects down to 1 to 2 microns. We recommend a method during scoping based on your container, your fill and the stage of your program.

What is the difference between deterministic and probabilistic methods?

Deterministic methods measure a physical property with a predictable, quantitative result, and can usually be validated to a defined detection limit. Probabilistic methods, such as dye ingress, depend on a series of events that each carry some uncertainty, so they need larger sample sizes and tighter control of test conditions.

How are positive controls made?

In your own container. We create laser-drilled defects of different sizes and locations in the glass, wire-channel defects at the stopper, and gross defects such as punctured or torn stoppers. Negative controls are units made by your normal process.

Can you test the syringe inside an autoinjector?

Yes. We develop and validate a disassembly method that frees the primary container from the device without disturbing the seals, then test the container.

What happens if our package changes?

A change to the container, closure, fill volume or stopper position can change what the method detects, so the method may need to be requalified. We will tell you during scoping which changes would trigger that.