ISO 11040 testing for prefilled syringes and cartridges
Component and finished-syringe testing for glass and polymer prefilled syringes, plunger stoppers and cartridges, used to choose a primary container and then to qualify it with your formulation.
What ISO 11040 covers
ISO 11040 is the series of standards for prefilled syringes. It describes the dimensional, functional and integrity characteristics of the barrel, the plunger stopper and the finished syringe.
Part 4 covers glass barrels and sterilized subassembled syringes ready for filling. Part 5 covers plunger stoppers, Part 6 covers plastic barrels, and Part 8 sets requirements and test methods for finished prefilled syringes. Supplier data usually describes the empty component. The filled syringe, with your formulation, your fill volume and your stopper position, is yours to characterize. We run all of the test annexes in the parts that apply to your container.
The syringe is also the most critical component in most autoinjectors, so the same data feeds device design. We use these methods in two ways: to compare candidate syringes, stoppers and cartridges before a supplier is chosen, and to follow the selected system with your formulation over its shelf life.
- Time zero
- After aging
What we test
Testing can be a single study on one question or a full characterization of each candidate container.
Dimensions and mechanics
Glass and polymer barrels, staked needle or luer, rigid and soft needle shields.
Critical dimensions
External geometry measured as the syringe rotates, referenced to the datums on your or your supplier's drawing. Barrel inner diameter is measured along the length at points 90 degrees apart, needle inner diameter is calculated from flow, and plungers are imaged for eccentricity.
Break-loose and glide force
Plunger break-loose and glide force in filled syringes, at the speeds you define, with your formulation or a matched surrogate, at time zero and at every aging timepoint.
Needle shield and needle tests
Needle shield removal force at the speed used for the supplier's lot release, needle penetration force and needle bond strength.
Strength and breakage
Flange breakage, tip and cone breakage, burst resistance and plunger stopper detachment, recorded as full force-displacement curves on fixtures and at speeds that follow the standard.
Lubrication, particles and cartridges
The parts of the system that change over time, and the cartridge formats used in pens and on-body injectors.
Silicone distribution
Silicone oil coverage is mapped over the full inner surface of the barrel before performance testing. The scan is non-destructive, so the same syringe can be followed through aging to see silicone migrate and move into solution. See custom siliconization.
Sub-visible particles
Flow imaging of the syringe contents gives count, size and shape for every sub-visible particle, so that silicone droplets can be told apart from other particles.
Integrity of the finished syringe
Container closure integrity of the stopper and tip or needle-shield seals, with leak test methods developed for your package. See container closure integrity testing.
Cartridges
Crimp-sealed cartridges for pens and on-body injectors: crimp and septum geometry, plunger forces, leakage, and septum fragmentation and resealability. Where the cartridge sits in a needle-based injection system, testing follows ISO 11608-3.
How we work
Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.
-
Scope
We review the candidate containers, your formulation and the questions you need answered, and propose a test matrix with sample quantities.
-
Develop
Methods and fixtures are set up for each container format. Filled-syringe methods are developed with placebo or surrogate wherever possible.
-
Validate
Methods that will support qualification or a filing are validated across operators and days, with a protocol, report and final method.
-
Test
Candidates are tested side by side at time zero and at each aging timepoint, with data released as each timepoint completes.
-
Report
You receive a report comparing candidates against your requirements, with raw data, run charts, box plots and normality checks.
Deliverables
Deliverables are defined in the proposal before work starts. A typical engagement includes the following.
Related services
Questions we hear often
If yours isn't here, it's a good thing to bring to a consultation.
What is the difference between ISO 11040-4 and ISO 11040-8?
Part 4 covers glass barrels and sterilized subassembled syringes that are ready for filling, so it is mainly about the component as supplied. Part 8 covers the finished prefilled syringe, filled and stoppered, so it is about the system with your formulation. Most programs need data on both.
Can you compare syringes and stoppers from several suppliers?
Yes. A common study is a matrix of two or more syringe options against two or more stopper options, each filled with your formulation, measured at time zero and followed over aging timepoints.
Do you test polymer syringes as well as glass?
Yes. We test glass and polymer barrels, with staked needles or luer connections, from any supplier.
Do you test cartridges?
Yes. We test crimp-sealed cartridges for pens and on-body injectors, including crimp and septum geometry, plunger forces, leakage, and septum fragmentation and resealability.
How many samples do you need?
It depends on the tests, the number of timepoints and the confidence level you need. The proposal lists sample quantities for each phase.
Choosing or qualifying a syringe?
Tell us which containers you are considering and what you need to know about them. We'll set up a call to talk through a test plan.
Prefer email? Write to info@kiln-lab.com or call +1 (866) 350-8240.