Home / Container closure integrity testing / Headspace analysis

Laser-based headspace analysis

Non-destructive, deterministic container closure integrity testing that reads the headspace of the sealed container, with a method developed and validated for your prefilled syringe or cartridge.

How headspace analysis works

Headspace analysis checks container closure integrity by measuring the gas above the product in the sealed container, without opening it.

Laser-based headspace gas analysis measures the gas composition or pressure in the headspace of the sealed unit. In an intact container those conditions are known. A breach in the container closure system shows as a change from the expected headspace conditions, and the unit is identified as a leak.

The method is non-destructive and deterministic: it measures a physical property and gives a quantitative result, which is why USP <1207> lists it among the leak test methods it describes. As with any CCI method, what it detects depends on the package. Container, closure, fill volume, headspace and stopper position all matter, so Kiln™ develops each method around your package and documents the detection limit during validation.

What we do

We confirm the method fits your package, make positive controls in your container, and develop and validate the method. Then we run it at the timepoints your program needs.

Method development and use

Each step is written into a protocol for your package before it is run.

Method selection

During scoping we review your container, closure, fill and headspace, and recommend headspace analysis only where it suits the package. If HVLD, helium leak, vacuum decay or dye ingress fits better, we say so.

  • Scoping
  • Method recommendation

Positive and negative controls

Positive controls are made in your own container: laser-drilled defects of varied size and location in the glass, wire-channel defects at the stopper, and gross defects such as punctured or torn stoppers. Negative controls come from your normal process.

  • Laser-drilled defects
  • Wire-channel defects

Development and validation

Parameters are optimized with placebo or surrogate wherever possible to save drug product. The method is validated per USP <1225> and USP <1207> under a protocol you approve, with test series run by different operators on different days. Negative controls are expected to pass and positive controls to fail.

  • USP <1225>
  • Detection limit

Containers inside devices

Where an autoinjector or needle safety device blocks access to the container, we develop and validate a disassembly method that frees the syringe or cartridge without disturbing the seals, then test the container.

  • Device disassembly
  • Autoinjectors

Stability, transit and air shipment

The validated method is applied at stability timepoints, after distribution simulation, and after simulated air shipment, where a headspace bubble expanding at altitude can push the stopper.

  • Stability
  • Post-transit CCI

How we work

Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.

  1. Scope

    We review the package, the product and the stage of your program, then agree the method, the fill and the sample plan.

  2. Develop

    Positive controls are made in your container, and method parameters are set using placebo or surrogate fills.

  3. Validate

    Test series by different operators on different days, to a protocol you approve. The detection limit is documented.

  4. Test

    Samples from development, stability, transit studies or production lots are tested against the validated method.

  5. Support

    The method stays available for later timepoints and investigations.

Deliverables

Deliverables are defined in the proposal before work starts. Sample sizes and detection limits are set for each program. A typical engagement includes the following.

Method recommendationWhether headspace analysis fits your package, and why
Positive controlsDefects of known size made in your own container and closure
Validation protocol and reportProtocol for your approval, then the report and data
Validated test methodA final method for stability, transit and investigation samples
Test reportsResults for each sample set, with the underlying data

Related services

Container closure integrity testingOverview of every CCI method we run, with validation to USP <1207>
High voltage leak detectionNon-destructive HVLD for liquid-filled syringes and cartridges
Helium leak testingHelium leak rate of each closure interface per ASTM F2391
Vacuum decay testingLeak detection from a pressure change in a chamber built for your container
Dye ingress testingProbabilistic testing where continuity with earlier data matters

Questions we hear often

If yours isn't here, it's a good thing to bring to a consultation.

What does headspace analysis measure?

Laser-based headspace gas analysis measures the gas composition or pressure in the headspace of the sealed container, without opening it. A leak shows as a change from the headspace conditions expected for an intact unit.

Is headspace analysis a deterministic method?

Yes. It measures a physical property and gives a quantitative result, and it is listed in USP <1207>. Deterministic methods can usually be validated to a defined detection limit, which is documented for your package during validation.

Is headspace analysis suitable for my syringe or cartridge?

It depends on the container, the closure, the fill and the headspace in the sealed unit. We review your package during scoping and recommend the method that fits it best, which may be headspace analysis, HVLD, helium leak, vacuum decay or dye ingress.

Do we have to use drug product?

Not for method development. Method parameters are optimized with placebo or surrogate wherever possible to save drug product, and we agree the fill for each phase during scoping.

Can you test a syringe or cartridge that is already in a device?

Yes. We develop and validate a disassembly method that frees the container from the autoinjector or safety device without disturbing the seals, then test the container.

Considering headspace analysis?

Tell us about the package and the product. We'll set up a call to talk through method fit, samples and timing.

Prefer email? Write to info@kiln-lab.com or call +1 (866) 350-8240.