Helium leak testing (ASTM F2391)
The inherent integrity of each closure interface of a prefilled syringe or cartridge, measured with an Agilent HLD MD30 helium mass spectrometer leak detector.
Why start with inherent integrity
Best practice is to study the inherent integrity of the empty container system first. It is the baseline that confirms the chosen container performs as designed, before the system is evaluated filled with product.
ASTM F2391 is the standard method for measuring package and seal integrity with helium as a tracer gas. Kiln™ isolates each closure interface with custom fixturing and measures its helium leak rate on an Agilent HLD MD30 helium mass spectrometer leak detector, so each result belongs to one seal rather than to the package as a whole. The HLD MD30 uses an oil-free dry scroll pump, so test parts are not exposed to pump oil, and has a rated sensitivity of 5 × 10-12 mbar·L/s (manufacturer specification). Results are analyzed against an acceptance limit set in the protocol with reference to USP <1207>.
Because the parts are tested empty, helium leak testing answers a different question from methods run on filled units. A typical program pairs the two: helium leak testing to establish inherent integrity, then a deterministic method such as high voltage leak detection on the filled system for stability, transit and lot testing.
What we test
Each seal interface is evaluated on its own, in your container and with your components.
Closure interfaces
The interfaces evaluated are agreed during scoping and written into the protocol.
Needle-side seal
On a prefilled syringe, the needle tip against the needle shield or tip cap, isolated from the rest of the package with custom fixturing.
Stopper-to-barrel seal
The plunger stopper against the barrel of a syringe or cartridge, tested separately from the other end of the container.
Cartridge crimp seal
On a cartridge, the crimp seal at the head of the container, isolated and measured as its own interface.
Component and configuration comparison
Because each interface is measured on its own, results can be compared across candidate components and configurations to support a component choice with data.
How we work
Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.
-
Scope
We review the container and its components, agree which interfaces to evaluate, and set the sample size for your program.
-
Feasibility
A feasibility study confirms the system is compatible with helium leak detection before the inherent integrity study starts.
-
Fixture
Custom fixturing is made to isolate each closure interface of your assembled package.
-
Test
Empty containers, sealed per your normal process parameters, are tested per ASTM F2391 against the acceptance limit in the protocol.
-
Report
You receive the test report and raw data, and a recommendation for evaluating the filled system next.
Deliverables
Deliverables are defined in the proposal before work starts. Sample sizes and acceptance limits are set for each program. A typical study includes the following.
Related services
Questions we hear often
If yours isn't here, it's a good thing to bring to a consultation.
Why are the containers tested empty?
The helium study establishes the inherent integrity of the container system as designed, before product is added. Samples are empty containers sealed per your normal process parameters. The filled system is then evaluated with a deterministic method such as HVLD.
Which seals do you test?
Each closure interface on its own. For a prefilled syringe that is usually the needle tip against the needle shield or tip cap, and the plunger stopper against the barrel. For a cartridge it is the stopper and the crimp seal. Each interface is isolated with custom fixturing.
What acceptance limit do you use?
Results are analyzed against an acceptance limit defined in the protocol with reference to USP <1207>. The limit and the sample size are agreed for each program before testing starts.
What if our container is not compatible with helium testing?
A feasibility study comes first to confirm that the system is compatible with helium leak detection. We only proceed with the inherent integrity study once compatibility is confirmed. If it is not, we review other methods with you.
Can helium testing compare components?
Yes. Because each interface is measured on its own, results can be compared across candidate components and configurations, such as different plunger stoppers or needle shields in the same container.