Home / Root cause investigations
Root cause investigations for combination products
When injection times drift, a test fails or a complaint comes in, we trace device performance back to the components, the fill and the assembly, and correlate the two with data.
Ending with a cause, not a guess
The symptom shows up at the device. The cause usually sits in a component, or in how the components meet the formulation.
Most combination product problems happen where the drug meets the device: a plunger that glides fine at T0 and sticks at twelve months, a seal that passes on paper and leaks after transit, a spring that can't push a viscous formulation through a thin-wall needle in time. A slow injection or a stalled stopper can come from barrel inner diameter, silicone level and distribution, the stopper, the needle, the fill or the assembly, and often from more than one of them.
Kiln™ characterizes the incoming components, tracks each one through fill, aging and testing, and correlates component variation to device performance. Where we qualified methods for your design verification or stability work, the investigation uses the same methods, so new results can be compared directly with the data you already have.
- 1Each point is one traced unit: a component measurement against its device result
- 2Trend: shows whether the component variation explains the performance
- 3A unit off the trend points to a second cause to follow up
What we look at
Each investigation is planned around the problem in front of you. Tests, sample sizes and the order of work are set per investigation.
Components
The syringe, cartridge, stopper and needle, measured before and after they go into the system.
Critical dimensions
Barrel inner diameter along its length, needle inner and outer diameter, needle length and bend, flange and plunger geometry, plunger position and headspace, and cartridge crimp and septum geometry. Measurement tools include a Keyence image dimension measurement system and an optical comparator.
Container mechanics
Break-loose and glide force, needle shield removal, breakage forces, needle bond strength and burst resistance, measured on Instron and Zwick universal test machines, with force-displacement curves reported with the values.
Silicone and particles
Silicone oil distribution is mapped across the barrel non-destructively, so the same syringe can be followed through aging. Sub-visible particles are imaged by FlowCam flow imaging to separate silicone droplets from protein aggregates. Drug product viscosity and density are measured where the formulation is part of the question.
Device and system
How the assembled product behaves, and where it stops behaving as intended.
Device function
Injection time and depth, activation and cap removal force, delivered dose, and break-loose and glide force of the container inside the device, run on qualified methods.
Container closure integrity
Suspect units are tested on a validated leak test method: high voltage leak detection on a PTI E-Scan 655, vacuum decay on a PTI vacuum decay tester, helium leak on an Agilent HLD MD30, headspace analysis or dye ingress, depending on the package. Where no validated method exists yet, one is developed for your package first.
Device disassembly
A defined disassembly method frees the syringe or cartridge from the pen, autoinjector or safety device. Each part is labeled for traceability, and for integrity testing the method is confirmed not to disturb the seals.
Transit and air shipment
When a problem appears after shipping, ASTM D4169 distribution sequences and simulated air shipment, with stopper position filmed throughout, help test whether transit is the cause.
Analysis and confirmation
Turning measurements into a cause you can act on.
Traceability and correlation
Every component is identified, measured and tracked through fill, aging and testing. Component variation is then correlated to device performance, with process capability analysis where the data support it.
Tolerance stack-up
Worst-case and statistical stack-ups of barrel, stopper and plunger rod drawings check whether a fit or function problem is possible within drawing tolerances, across the supplier combinations in use.
Testing at the extremes
To test a suspected cause directly, units can be built at its extremes: syringes re-siliconized to the low and high ends of a supplier's specification, fills at nominal and process extremes, and assemblies with force and torque monitored on every unit.
How we work
Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.
-
Scope
We review the failure or complaint, your existing data and the units and components available, and propose a plan, tests and sample requirements.
-
Characterize
Components and suspect units are measured: dimensions, forces, silicone, particles and integrity, on methods qualified or validated for your system.
-
Trace
Each component is labeled and followed through fill, assembly, aging and testing, so every result links back to the parts that produced it.
-
Correlate
Component variation is compared with device performance, and stack-ups or tests at the extremes check the suspected cause.
-
Report
You receive the findings with the full data package. The methods stay available for incoming inspection and lot release testing.
Deliverables
Deliverables are defined in the proposal before work starts. A typical investigation includes the following.
Related services
Questions we hear often
If yours isn't here, it's a good thing to bring to a consultation.
What kinds of problems do you investigate?
Injection times that drift, forces that change with aging, unexpected failures in design verification or stability testing, leaks found after transit, and complaints. We look for the cause in the components, the fill and the assembly, and in how they interact with the formulation.
Can you use the methods from our earlier testing?
Yes. If we qualified the methods for your design verification or stability work, the investigation uses the same methods, so new results compare directly with earlier data. Methods from your own lab can be brought in by method transfer.
Do you need assembled devices or loose components?
Both are useful. Loose components can be characterized before assembly and tracked from there. Assembled devices are taken apart by a defined disassembly method that frees the syringe or cartridge, and each part is labeled for traceability.
How do you confirm a suspected cause?
Where possible, by testing it directly. Units can be built at the extremes of the suspected variable, for example syringes re-siliconized to the low and high ends of a supplier's specification or fills at process extremes, and then tested on the same methods. Tolerance stack-ups check whether a fit problem is possible across the component combinations in use.
How many samples do you need?
It depends on the problem, the components involved and the tests needed to separate one possible cause from another. We set sample requirements per investigation and spell them out in the proposal.
Investigating a failure or a complaint?
Tell us what you are seeing, what data you have and which units and components are available. We'll set up a call to talk through a plan, samples and timing.
Prefer email? Write to info@kiln-lab.com or call +1 (866) 350-8240.