Home / Functional testing
Functional testing for autoinjectors, pens and syringes
Device and primary container methods developed and qualified on your system: forces, dose accuracy, injection time and depth, through conditioning, aging and transit.
Testing the system, not just the parts
A combination product works only if the device and the primary container work together, at every condition the product will see.
Functional testing measures what the user and the patient depend on: the force to remove the cap and activate the device, the time to deliver the dose, how deep the needle goes and how much drug comes out. For the primary container it covers the force to start the plunger stopper moving and keep it moving, needle shield removal and breakage forces. ISO 11608-1 sets the requirements for needle-based injection systems, ISO 11608-5 for automated functions and ISO 11040-8 for finished prefilled syringes. We run all test annexes of the parts that apply to your system.
A result at T0 is only the start. A plunger that glides fine at T0 can stick at twelve months, and a spring has to push the formulation through the needle in time at every condition. Kiln™ develops and qualifies each method on your system first, then uses the same method at every aging timepoint, after conditioning and after transit, so results can be compared across the program.
- 1Break-loose force: the peak needed to start the plunger stopper moving
- 2Glide force: average and maximum force while the stopper travels
- 3End of travel: total displacement, recorded with the forces
What we test
Methods are written for your device and container, not pulled off a shelf. Sample sizes, conditions and timepoints are set per program.
Device function
What it takes to use the device, and what it delivers.
Device forces
Cap removal, activation and override force, lock-out force, and needle shielding and separation force, measured on Instron and Zwick universal test machines with fixtures made for your device. Torque measurements run on an Instron torsion tester.
Injection time and depth
Injection time, injection depth and click-to-click time, measured on the actuated device with methods developed for your design.
Dose accuracy
Delivered dose and deliverable volume are measured on Mettler Toledo analytical balances, for single-dose devices and multi-dose pens. Graduation and dose-mark accuracy are covered where the device has them.
Pens, cartridges and needles
Septum fragmentation and resealability per ISO 11608-3, cartridge and plunger leakage, and needle leakage.
Primary container
The syringe or cartridge, tested on its own or freed from the device.
Break-loose and glide force
Break-loose force is the force needed to start the plunger stopper moving. Glide force is the force to keep it moving along the barrel. We measure both per ISO 11040-8 and report break-loose force, average and maximum glide force and total displacement. Where the container sits inside a pen or autoinjector, we free it by a defined disassembly method first.
Container mechanics
Needle shield removal, flange breakage, tip and cone breakage, needle penetration force, needle bond strength and plunger stopper detachment, with fixtures made for your container.
Conditioning, aging and transit
The same qualified methods, applied after everything storage and shipping can do to the product.
ISO 11608-1 series testing
Functional and dose accuracy testing after preconditioning at cool, standard and warm atmospheres, after dry heat and cold storage, after free fall in three orientations, and after vibration in three axes per IEC 60068-2-6. Conditions are confirmed against the current edition of the standard and your device's designation.
Aging and functional stability
Real-time aging per ICH Q1A(R2) at your storage condition, including 2–8 °C, and accelerated aging per ASTM F1980, with functional testing at every timepoint. Time out of refrigeration can be added before testing where your program needs it.
Simulated distribution
Full shippers go through ASTM D4169 sequences: manual handling, vehicle stacking, loose-load and vehicle vibration, low pressure and, where required, concentrated impact. The devices then go on to visual inspection, functional and dose accuracy testing. Simulated altitude testing, for example for air shipment, runs in a FlexPak Package Leak Detector vacuum chamber.
How we work
Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.
-
Scope
We review the device, the container, your design inputs and the standards that apply, and propose methods, conditions, timepoints and sample requirements.
-
Develop
Fixtures and methods are developed on your system, using placebo or surrogate fills wherever possible to save drug product.
-
Qualify
On your approval, each method is qualified to a protocol, and you receive the qualification report and the qualified method.
-
Execute
Design verification, ISO 11608-1 series, aging timepoints and transit studies run on the qualified methods, with a report at each milestone.
-
Support
Qualified methods stay available for later timepoints, investigations and complaints, or transfer to your lab with fixtures.
Deliverables
Deliverables are defined in the proposal before work starts. A typical engagement includes the following.
Related services
Questions we hear often
If yours isn't here, it's a good thing to bring to a consultation.
Which standards do you test to?
ISO 11608-1 for needle-based injection systems, ISO 11608-3 for containers and integrated fluid paths, ISO 11608-5 for automated functions, and ISO 11040-8 for finished prefilled syringes. We run all test annexes of the parts that apply to your system, alongside your own design inputs.
How are devices conditioned before testing?
As ISO 11608-1 requires, devices are preconditioned at cool, standard and warm atmospheres, and exposed to dry heat and cold storage, before functional and dose accuracy testing. We confirm the conditions against the current edition of the standard and the storage limits in your instructions for use.
Can you test the syringe or cartridge inside a pen or autoinjector?
Yes. We disassemble the device by a defined method to free the container, label each part for traceability, and then test break-loose and glide force or other container properties.
Do the methods carry over to stability and investigations?
Yes. Each method is qualified once. The qualified method is then used at aging timepoints, before and after distribution testing, and in failure investigations, so results stay comparable across the program.
How many samples do you need?
It depends on the tests, conditions and timepoints, and on your design verification sampling plan. We set sample requirements per program and spell them out in the proposal.
Need functional data on your device?
Tell us about the device, the container and where you are in the program. We'll set up a call to talk through methods, samples and timing.
Prefer email? Write to info@kiln-lab.com or call +1 (866) 350-8240.