Dye ingress testing

A probabilistic container closure integrity test for prefilled syringes and cartridges, run with defined challenge conditions and positive controls made in your container.

Where dye ingress fits

Test units are immersed in a blue dye solution with surfactant and put through a vacuum challenge in a FlexPak Package Leak Detector vacuum chamber. A leak shows as blue dye inside the container, found by visual inspection.

Dye ingress is a probabilistic method. Each result depends on a series of events that each carry some uncertainty, so the method needs larger sample sizes and tighter control of test conditions than a deterministic one. USP <1207> favors deterministic methods, which measure a physical property and give a quantitative, repeatable result.

For that reason we usually recommend a deterministic method first. Kiln™ runs dye ingress where a program needs continuity with existing dye ingress data. As with every CCI method, it is built around your package: fill volume, headspace, stopper position, the product and each seal interface all change what the method can detect.

What we do

We confirm the method choice with you, build positive controls in your container, and develop and validate the method. Then we run it at the timepoints your program needs.

Method development and use

Each step is written into a protocol for your package before it is run.

Method selection

During scoping we review the package, the product and your existing data, and confirm whether dye ingress or a deterministic method fits the question your program needs answered.

  • Probabilistic
  • Method recommendation

Positive and negative controls

Positive controls are made in your own container: laser-drilled defects of varied size and location in the glass, wire-channel defects at the stopper, and gross defects such as torn or punctured stoppers. Negative controls come from normal processing.

  • Laser-drilled defects
  • Wire-channel defects

Method development

Units are immersed in blue dye with surfactant in a FlexPak Package Leak Detector vacuum chamber, and the vacuum challenge is defined for your package and held under tight control. Each unit is then inspected visually for dye inside the container. The same FlexPak Package Leak Detector vacuum chamber is also used for bubble emission and simulated altitude testing. Method parameters are optimized with placebo or surrogate where possible to save drug product, and sample sizes are set for your program.

  • Vacuum challenge
  • Visual inspection
  • Placebo or surrogate

Validation

The method is validated under a protocol you approve before execution, with acceptance criteria set in the protocol: negative controls are expected to pass and positive controls to fail. You receive the protocol, the report and a final test method.

  • USP <1225>
  • Protocol and report

Containers inside devices

Where a device blocks access to the container, we develop and validate a disassembly method that frees the syringe or cartridge without disturbing the seals, then test the container.

  • Device disassembly
  • Autoinjectors

Stability and transit

The validated method is applied at stability timepoints and after distribution simulation, so new results can be read alongside your earlier dye ingress data.

  • Stability
  • Post-transit CCI

How we work

Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.

  1. Scope

    We review the package, the product and your existing data, confirm the method, and agree the fill and sample plan for each phase.

  2. Develop

    Positive controls are made in your container, and challenge conditions and method parameters are set with placebo or surrogate.

  3. Validate

    Test series run to a protocol you approve, against acceptance criteria defined before execution. The result is a final test method.

  4. Test

    Samples from development, stability, transit studies or production lots are tested against the validated method.

  5. Support

    The method stays available for later timepoints and investigations, or transfers to your lab.

Deliverables

Deliverables are defined in the proposal before work starts. Sample sizes are set for each program. A typical engagement includes the following.

Method recommendationWhether dye ingress or a deterministic method fits your package, and why
Positive controlsDefects of known size made in your own container and closure
Validated methodValidation protocol, report and final test method
Test reportsResults for each sample set, with the underlying data

Related services

Container closure integrity testingAll CCI methods, positive controls and validation to USP <1207>
High voltage leak detectionNon-destructive testing of liquid-filled syringes and cartridges
Helium leak testingInherent integrity of each closure interface per ASTM F2391
Headspace analysisLaser-based headspace gas analysis of the sealed container
Vacuum decay testingNon-destructive leak testing in a chamber made for your container

Questions we hear often

If yours isn't here, it's a good thing to bring to a consultation.

How is the dye ingress test run?

Units are immersed in a blue dye solution with surfactant in a FlexPak Package Leak Detector vacuum chamber and put through a vacuum challenge defined for your package. After the challenge, each unit is inspected visually for blue dye inside the container. The challenge conditions and acceptance criteria are set in the protocol.

What else is the FlexPak vacuum chamber used for?

Besides dye ingress, we use the FlexPak Package Leak Detector vacuum chamber for bubble emission leak testing and for simulated altitude testing, such as checking stopper movement during simulated air shipment.

When is dye ingress the right choice?

Mainly when your program needs continuity with existing dye ingress data. USP <1207> favors deterministic methods, so we usually recommend HVLD, vacuum decay, headspace analysis or helium leak first. We agree the method during scoping based on your container, your fill and the stage of your program.

Why does dye ingress need more samples than a deterministic method?

Dye ingress is probabilistic: each result depends on a series of events that each carry some uncertainty, so the method needs larger sample sizes and tighter control of test conditions. Sample sizes are set for each program and agreed in the proposal.

How are positive controls made?

In your own container. We create laser-drilled defects of different sizes and locations in the glass, wire-channel defects at the stopper, and gross defects such as punctured or torn stoppers. Negative controls are units made by your normal process.

Do we have to use drug product?

Not for method development, where it can be avoided. Method parameters are optimized with placebo or surrogate solution to save drug product, and we agree the fill for each phase during scoping.

Can you test a syringe that is already assembled into a device?

Yes. We develop and validate a disassembly method that frees the syringe or cartridge from the device without disturbing the seals, then test the container.

Need a dye ingress method?

Tell us about the package, the product and the data you already have. We'll set up a call to talk through method options, samples and timing.

Prefer email? Write to info@kiln-lab.com or call +1 (866) 350-8240.