High voltage leak detection (HVLD)

Non-destructive container closure integrity testing for liquid-filled prefilled syringes and cartridges, with a method developed and validated for your package.

How HVLD works

High voltage leak detection, also called alternating-direct high voltage (ADHV) leak detection, finds breaches in non-porous, rigid and non-rigid packages that hold a liquid product.

The filled unit rotates on its horizontal axis while electrical probes scan its outer contours, and the conductivity signal is measured throughout the test cycle. A breach in the container closure system shows as a resistance differential and a change in current flow. HVLD can detect defects down to 1 to 2 microns.

The test is non-destructive, highly repeatable and needs little operator handling. Once validated, the same method can be used at many stages of the product life cycle: functional stability, failure investigations, and testing before and after distribution stresses. Kiln™ runs HVLD on a PTI E-Scan 655 and builds each method around the specific package, so the limit of detection is established for your container, closure and fill.

What we do

We confirm feasibility, build positive controls in your container, and develop and validate the method. Then we run it at the timepoints your program needs.

Method development and validation

Each step is written into a protocol for your package before it is run.

Feasibility

A first study confirms that HVLD suits your package and fill before method development starts. You receive a feasibility report.

  • Feasibility report

Positive and negative controls

Positive controls are laser-drilled defects in the container wall, varied in size and location, plus gross defects such as holes or tears in the plunger stopper or needle shield. Before testing, each positive control is prepared so the defect is filled with liquid. Negative controls are units processed per your normal parameters.

  • Laser-drilled defects
  • Gross defects

Method development

Test samples can be assembled at Kiln™ and filled with placebo, surrogate solution or drug product. Development uses placebo or surrogate where possible to save drug product. We set test parameter limits and pass/fail limits, run trials with positive controls to establish the limit of detection, and verify repeatability, reliability and robustness.

  • Pass/fail limits
  • Limit of detection

Validation

The method is validated per USP <1225> and USP <1207> under a protocol you approve before execution. Test series are run by different operators on different days to show intermediate precision, against acceptance criteria set in the protocol: negative controls are expected to pass and positive controls to fail. You receive a final test method.

  • USP <1225>
  • Intermediate precision

Containers inside devices

Where a needle safety device or autoinjector blocks access to the container, we develop and validate a disassembly method that frees the syringe or cartridge without disturbing the seals, then test it with HVLD.

  • Device disassembly
  • Needle safety devices

Records and deviations

Component lots and the instruments used, with their calibration history, are recorded in the raw data. Any protocol amendment is agreed with you in advance, and any deviation is documented in the final report.

  • Traceability
  • Documented deviations

How we work

Work is structured in phases with clear deliverables at each gate, so you always know what you have paid for and what comes next.

  1. Scope

    We review the package, the product and the stage of your program, and agree the fill and sample plan for each phase.

  2. Develop

    Feasibility is confirmed, positive controls are made in your container, and pass/fail limits and the limit of detection are established.

  3. Validate

    Test series by different operators on different days, run to a protocol you approve. The result is a final test method.

  4. Test

    Samples from stability, distribution studies, investigations or production lots are tested against the validated method.

  5. Support

    The method stays available for later timepoints and investigations, or transfers to your lab.

Deliverables

Deliverables are defined in the proposal before work starts. Sample sizes and detection limits are set for each program. A typical engagement includes the following.

Feasibility reportWhether HVLD suits your package and fill
Method development reportParameters, pass/fail limits and limit of detection, with the data package
Validation protocol and reportProtocol for your approval, then the report and data package
Validated test methodA final method for use across the product life cycle
Test reportsResults for each sample set, with the underlying data

Related services

Container closure integrity testingAll CCI methods, positive controls and validation to USP <1207>
Helium leak testingInherent integrity of each closure interface per ASTM F2391
Headspace analysisLaser-based headspace gas analysis of the sealed container
Vacuum decay testingNon-destructive leak testing in a chamber made for your container
Dye ingress testingA probabilistic method for continuity with existing data

Questions we hear often

If yours isn't here, it's a good thing to bring to a consultation.

Is HVLD suitable for my product?

HVLD applies to non-porous, rigid and non-rigid packages that hold a liquid product, such as prefilled syringes and cartridges. We confirm it works for your package and fill with a feasibility study before method development starts.

How small a defect can HVLD detect?

HVLD can detect defects down to 1 to 2 microns. The limit of detection for your package is the smallest nominal flow effective diameter that is reliably detected in trials with positive controls, and it is documented in the method development and validation reports.

Does HVLD damage the units it tests?

No. The test is non-destructive and needs little operator handling, which makes it practical for stability timepoints, failure investigations and testing before and after distribution.

Do we have to use drug product?

Not for method development. Methods are developed with units filled with placebo or surrogate solution to save drug product. Units filled with drug product can be used where your program needs it, and we agree the fill for each phase during scoping.

Can you test a syringe that is already assembled into a device?

Yes. We develop and validate a disassembly method that frees the syringe or cartridge from the needle safety device or autoinjector without disturbing the seals, then test the container with HVLD.